|
Only the used device was returned loose inside the carton.The lens stop and plunger lock were removed.The plunger was oriented correctly.Inadequate viscoelastic was observed in the device.The plunger was retracted to mid-nozzle.The nozzle tip beveled edge was slightly bent backward.The lens was not returned.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.A non-qualified viscoelastic was indicated.The root cause for the reported complaint was most likely related to a failure to follow the instructions for use (ifu).An inadequate amount of viscoelastic was observed and a non-qualified viscoelastic was indicated.The ifu instructs: fully insert the ophthalmic viscosurgical device (ovd) cannula, perpendicular to the device, through the viscoelastic port, located in the lens stop portion of the device.Fill the device until ovd can be observed flowing to the ¿fill-to¿ line on the nozzle tip.This will require approximately 0.2 milliletre of ovd.Only use an company ovd qualified for use with the company pre-loaded delivery system that has been allowed to come to the operating room temperature.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device.The ifu instructs: during device preparation and implantation of the company intraocular lens (iol) with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.The ifu instructs: take at least 7 seconds to gently fold the iol by advancing the plunger forward in one smooth, continuous motion until the front edge of the optic is even with the ¿fill-to¿ line on the nozzle.Do not allow any portion of the lens to exit the nozzle tip.At this position the blue spring will contact the clear main body.It is important to not advance the plunger abruptly to fold the iol as improper folding and lens damage may occur.The ifu instructs: after the lens has been advanced to the nozzle line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.After confirming the lens is properly positioned and the haptics are folded properly, proceed with lens implantation.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a broken haptic or other negative outcome, since the haptic may be trapped and stretched, and/or pinched and sheared by the moving plunger.The trailing haptic may become stuck: if the plunger is not fully advanced the haptic may not release properly from the device.Due to the use of a non-qualified viscoelastic.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.If the operating room temperature is too high (more than 23 degrees celsius or 73 degrees fahrenheit) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.The manufacturer internal reference number is: (b)(4).
|