| Model Number |
LUCAS 3 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device alarmed and paused.In this state, a delay in cardiopulmonary resuscitation may occur.If a device malfunctions, the device operating instructions indicate that the user is to remove the device and immediately start manual chest compressions.There was no report of patient harm associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker contacted the customer to request additional information on the patient.Stryker requests patient information necessary for regulatory compliance, strictly adhering to hipaa's privacy rule.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.A stryker service representative performed an evaluation of the customer¿s device and was unable to verify and duplicate the reported issue.After completing some unrelated repairs, proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.
|
| |
|