Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional did not report an adverse event.Patient later reported "rupture".This record is for the left side.The device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional did not report an adverse event.Patient later reported "rupture".This record is for the left side.The device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional did not report an adverse event.Patient later reported "rupture".Healthcare professional later reported "leaking".This record is for the left side.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: a4, b6, d4, g2.
|
| |
|
Event or Problem Description
|
|
Healthcare professional did not report an adverse event.Patient later reported "rupture".Healthcare professional later reported "leaking".This record is for the left side.The device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional did not report an adverse event.Patient later reported "rupture".Healthcare professional later reported "leaking".This record is for the left side.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- rupture: observed broken device through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (non-penetrating nick) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional/corrected data: d3, d6b.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6b, h.6.
|
| |
|