| Catalog Number |
101/870/070 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that an emergency trach change was performed due to a cuff issue.The cuff popped off without much pressure.There was patient involvement, no injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6.Codes: updated.No product or photos were returned therefore no device analysis could be completed.A device history record (dhr) review showed there were no discrepancies or non-conformances during the manufacturing of the reported lot number.If the product is returned the manufacturer will re-open the complaint for further device analysis.
|
| |
|