• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS Back to Search Results
Model Number DCB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vitreous Loss (2142)
Date of Event 03/02/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Section a2, a4 and a5: unknown/ not provided.Patient information cannot be provided due to personal data privacy legislation/policy.Section d6b: explant date: not applicable, as lens was not explanted.Section e1: initial reporter first & last name: information unknown/not provided.Section e1: email address: unknown/not provided.Section e1: telephone number: unknown, information not provided.Device evaluation: the product testing could not be performed as the product was not returned (the lens remains implanted).The reported complaint cannot be confirmed.Manufacturing record evaluation: based on the manufacturing records review, and historical complaint review, there is no indication of a product malfunction or product quality deficiency.The manufacturing records review for this production order (po) shows that the units were released within specification.No nc/er was found as part of this manufacturing records review.For complaint history review, a search of complaints related to this production order (po) was performed in the system.The search revealed that no other complaints were received for this po.No nonconformity report, documentation or labeling changes, and escalations are required.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that after implanting intraocular lens (iol) with model dcboo and during irrigation and aspiration(viscoelastic removal), the optic was not centered to the capsule bag and vitreous prolapse occurred.Anterior vitrectomy performed.No further information available.This report captures information for iol.Another report will capture information for phaco system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key24722658
Report Number3012236936-2026-000078
Device Sequence Number10204467
Product Code HQL
UDI-Device Identifier05050474636118
UDI-Public(01)05050474636118(17)280711
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDCB00
Device Catalogue NumberDCB0000215
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2026
Initial Report FDA Received Date03/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-