• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WAVELIGHT GMBH WAVELIGHT FS200 FEMTOSECOND LASER; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WAVELIGHT GMBH WAVELIGHT FS200 FEMTOSECOND LASER; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Catalog Number 8065990941
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Keratitis (1944)
Date of Event 01/02/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H.3., h.6.: investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that the patient experienced diffuse lamellar keratitis in the right eye after refractive surgery.The patient¿s symptoms subsequently improved.The procedures were performed by two experienced surgeons.The hospital confirmed that there have been no recent changes in postoperative medication practices, and no changes to surgical instruments.There are multiple related reports for this facility.This report addresses the patient initial nz in the right eye and other manufacturer reports will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVELIGHT FS200 FEMTOSECOND LASER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
Manufacturer (Section G)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM   91058
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24723092
Report Number3003288808-2026-00177
Device Sequence Number16035244
Product Code GEX
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K101006
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number8065990941
Device Lot Number1417X3
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2026
Supplement Date Received by Manufacturer07/24/2026
Initial Report FDA Received Date03/30/2026
Supplement Report FDA Received Date08/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
WAVELIGHT FS200 EASYPACK PATIENT INTERFACE
Outcome Attributed to Adverse Event Other;
Patient Age23 YR
Patient SexFemale
-
-