MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507325BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent primary breast augmentation with two 325cc mentor memorygel breast implants.Post-operatively, the patient was diagnosed, via physical examination, with bilateral breast capsular contractures (baker grades iv).As a result, the patient underwent breast implant removal and replacement surgery on (b)(6) 2026.During the removal surgery, both implants were found to be ruptured.The replacement devices were two new mentor gel implants.This medwatch form is for the left breast prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.The mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture, material rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On april 1, 2026, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted visual inspection of the returned device.Visual analysis of the returned device revealed that the breast implant was found to have a tear expanding from the posterior to the anterior aspect, measuring approximately 8.9 cm.Additionally, it had shell abrasion on the posterior edge of the rent.These findings are consistent with normal wear of the device.Shell abrasion suggests in-vivo folding or creasing of the device, which may be caused by continuous and sustained stresses to the device, such as an excessively small breast pocket or the folding or wrinkling of the shell within the breast pocket.In some cases, breast implants may also experience normal wear over time.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.
|
| |
|