Catalog Number ROB30001
Medical Device Problem Code Application Program Problem (2880)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, before a cori assisted tsa, the pre-op plan was uploaded to real intelligence cori console xt, all of the patient information loaded in and presumed to look completely fine.When pressing begin operation, an error was encountered that stated the pre-op plan file cannot be accessed.The case was deleted and uploaded multiple times.Encountered the same error every time.The procedure was performed, without any delay, with a change in surgical technique to manual procedure.No further complications were reported.
 
Additional Manufacturer Narrative
H11: internal complaint reference (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H6: results of investigation: the real intelligence cori console xt, part number rob30001, serial number (b)(6), was not returned for evaluation.An image was provided.The reported problem was confirmed with a visual inspection.The device showed a preop plan error.A complaint history review was performed by part number, and no similar complaints were identified.A review by lot/serial number was performed and only 1 similar complaint was identified.A review of manufacturing records indicate the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file, and the documented risk level is still adequate.A historical review concluded that no prior escalation actions are applicable to the scope of the reported complaint.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Although the reported problem was confirmed through a visual inspection, a definitive cause could not be determined.Factors contributing to the reported problem may have been associated with issues during collection/registration.Based on the investigation, no containment or corrective action is recommended or required at this time.Should any additional information be received, the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE CORI CONSOLE XT
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24724159
Report Number3010266064-2026-00058
Device Sequence Number10360641
Product Code OLO
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K252841
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberROB30001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2026
Supplement Date Received by Manufacturer04/30/2026
Initial Report FDA Received Date03/30/2026
Supplement Report FDA Received Date05/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-