|
It was reported that during a laparoscopic procedure, the device partially fired, became lodged on the specimen, and became stuck, r equiring removal by cutting it out of the patient¿s body.These complications led to an extended surgical time.The issue was resolved by resecting and re-stapling using a new reload, pli40 medtronic's initial evaluation of the incident device found that several millimeters of material were found to be sheared from the distal end of the firing rod.Pli50: medtronic's initial evaluation of the incident device found, the incomplete retraction could be traced to the damage to the firing rod of returned adapter.Pli60: medtronic's initial evaluation of the incident device found, the incomplete retraction could be traced to the damage to the firing rod of returned adapter.Pli70: medtronic's initial evaluation of the incident device found, the incomplete retraction could be traced to the damage to the firing rod of returned adapter.
|
|
D10 concomitant products : egia60amt - egia60amt egia 60 artic med thick sulu - lot# unknown sigphandle - sig power sigphandle handle - sn: unknown sigadaptstnd - sig power sigadaptstnd linear adapter - sn: (b)(6) sigadaptshort - sig power sigadaptshort lin short adapt - sn: (b)(6) sig60amt - tri-staple 2.0 60 artic med thick sulu - lot# p1m0570 sig30ctav - tri 2.0 sig30ctav 30 ct vsclr 12mm - lot# n4e1903y medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|