MAUDE Adverse Event Report: EXACTECH, INC. OPTETRAK ASY,CR CEMENTED FEMORAL, SZ 2,; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. OPTETRAK ASY,CR CEMENTED FEMORAL, SZ 2,; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 230-02-02
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Metal Related Pathology (4530); Swelling/ Edema (4577)
Date of Event 02/26/2026
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately 9 years and 11 months post the initial left total knee arthroplasty, the patient complained of knee swelling and pain.The surgeon diagnosed wear of the tibial insert and the patient was revised.Breakage and wear were observed in the retrieved tibial insert.The surgeon decided there was no problem with the locking mechanism of the tibial tray, and only the insert was replaced with a new truliant cr size 2, 13mm neutral insert.Polyethylene particles were removed by the surgeon appropriately.The surgeon found metallosis, but the tibial tray was left.There was no reported surgical delay/prolongation.The patient was last known to be in stable condition following the event.
 
Additional Manufacturer Narrative
The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTETRAK ASY,CR CEMENTED FEMORAL, SZ 2,
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
ashley harris
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key24727176
Report Number1038671-2026-00347
Device Sequence Number16000317
Product Code JWH
UDI-Device Identifier10885862061676
UDI-Public10885862061676
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K032606
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/09/2020
Device Catalogue Number230-02-02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2026
Initial Report FDA Received Date03/30/2026
Date Device Manufactured08/11/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age88 YR
Patient SexFemale
Patient Weight48 KG
-
-