The device was returned to olympus for inspection where a residue foreign material was confirmed.Based on the results of the investigation, olympus confirmed foreign material came off of the device, but the type of the material or root cause cannot be identified.A definitive root cause cannot be determined.According to previous investigations, the use of products that contain silicone can cause deposits of white foreign material.Silicone is for example included in simethicone, a defoaming agent, lubricants and so on.If the customer used them, there was risk that foreign material remained in the channel even if the reprocessing was conducted in accordance with ifu.Simethicone and petroleum/oil/silicone-based lubricants are non-water soluble and thus not recommended for use by olympus.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|