| Model Number |
133650 |
| Medical Device Problem Codes |
Difficult to Open or Close (2921); Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an open procedure, the clip applier had jaws that were not able to close when the handle was squeezed and there was an issue with loading or firing of the clips, where some clips could load and some could be fired.Another clip applier was used to complete the case.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the instrument was partially applied with nineteen remaining clips.The distal end of the channel cover was intact.Microscopic inspection of the jaws revealed acceptable jaw co-planarity.Functional testing noted that the instrument was cycled, and the pusher did not load the clip.The pusher bar was observed to be damaged.It was reported that the clips were not loading properly.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when the jaws of the instrument are constrained or with a side load to the jaws while attempting to fire a clip.It was also reported that the jaws would not close.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: do not excessively twist or torque the jaws.Excessive twisting or torquing the jaws may dislodge clip from the jaws or cause clip m alformation after jaw closure.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|