| Model Number |
GIF-2TH180 |
| Medical Device Problem Codes |
Erratic or Intermittent Display (1182); No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the gastrointestinal videoscope had no image and only static was displayed.The issue occurred during device setup.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, h2, h3, h6, h11 the device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the complaint was traced to corrosion found on the electrical connector.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|
Search Alerts/Recalls
|