|
It was reported that hole in the device occurred.The 100% stenosed target lesion was located in the mildly tortuous and moderately calcified superficial femoral artery.A 35, 90cm rubicon 35 was selected for use.During the procedure.The device became kinked, and the guidewire exited through a hole at the kinked segment.No troubleshooting steps were performed, and the device was replaced with a new catheter to complete the procedure.No patient complications were reported, and the patient fully recovered as a result of this event.
|