| Model Number |
TRB24-REG |
| Medical Device Problem Code |
Unintended Deflation (4061)
|
| Health Effect - Clinical Code |
Hematoma (1884)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Implanted date: device was not implanted.D6b: explanted date: device was not explanted.E3: occupation: territory manager.The actual device has not been returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.
|
| |
|
Event or Problem Description
|
|
Terumo medical corporation received the following reported information: following a left heart cath, the tr band was applied.However, a deflation issue with the balloon occurred and led to a hematoma.The estimated blood loss was less than 250cc.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.The complaint cannot be confirmed for air leakage issues because a sample was not returned for assessment.Without a sample, the exact root cause cannot be determined.Review of the device history record showed there were no issues noted during the manufacture of the product that could have contributed to this complaint condition.Currently, no action is recommended since the risk evaluation is within the predetermined limits in the hazard based risk table (hbrt).
|
| |
|