| Model Number |
VTICM5 13.2 (-13.00/2.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/18/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.H6 - investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault and elevated iop.Lens explanted.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.Cause of the event was reported as user error - wtw was not accurate.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation - lens was returned in a microcentrifuge tube in liquid.Lens returned in pieces.Unable to perform dimensional inspection due to significant lens damage.Claim# (b)(4).
|
| |
|