| Catalog Number |
00515047500 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).The device was returned with the tubing removed from the handpiece.Functional testing found the device would not power on with the original battery pack but did power on in both modes when connected to a lab battery pack.Visual inspection of the interior of the pack found the batteries had leaked electrolyte inside of the pack.The batteries were installed in the correct orientation inside the pack.There did not appear to be damage to the wiring of the pack.No other damage was found.Review of the device history record identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to manufacturing and design issues.The reported event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.E1 telephone number: (b)(6).G2 country: japan.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery on (b)(6) 2025, the device did not work at all and the handpiece and battery were getting hot.There was no patient impact but it is unknown if there was a delay.An alternate device was utilized to complete the procedure.During product evaluation it was found that the battery leaked electrolyte.Due diligence is complete and there is no additional information available.
|
| |
|