• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. CEMENT RESTRICTOR DIA.24MM; MESH, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. CEMENT RESTRICTOR DIA.24MM; MESH, SURGICAL Back to Search Results
Catalog Number EBO101
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 03/02/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information has been requested and if received, will be provided in the supplemental report.
 
Event or Problem Description
It was reported that, a removal of latitude ev total elbow replacement was performed for a patient who had an infection.The patient had a revision surgery on (b)(6) 2025.The surgeon accessed the joint and removed a significant amount of black matter and then proceeded to take the humeral and ulnar stem out.No further information was provided.
 
Event or Problem Description
It was reported that, a removal of latitude ev total elbow replacement was performed for a patient who had an infection.The patient had a revision surgery on (b)(6) 2025.The surgeon accessed the joint and removed a significant amount of black matter and then proceeded to take the humeral and ulnar stem out.No further information was provided.Later it was added that after speaking to the surgeon he mentioned the patient had gone well and was happy with the outcome of plating the patient's arm with another companies products.
 
Additional Manufacturer Narrative
Correction-date of explant (d6b) added.The reported event could not be confirmed since the device was not returned for evaluation and no other evidence besides a photograph of the explanted devices were provided.The device inspection revealed the following.An evaluation of the provided images shows the devices to be intact, with all components account for and a complete cement mantle present.Without any medical imaging of the alleged event and no pathology provided, the event cannot be confirmed.A functional inspection was not possible as the device was not returned.A dimensional inspection was not possible since the devices were not returned.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CEMENT RESTRICTOR DIA.24MM
Common Device Name
MESH, SURGICAL
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR   38330
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key24733514
Report Number3000931034-2026-00319
Device Sequence Number9242028
Product Code FTM
UDI-Device Identifier03700386922053
UDI-Public(01)03700386922053(11)250308(17)300308(21)5804BB024
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K973453
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date03/08/2026
Device Catalogue NumberEBO101
Device Lot Number5804BB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/02/2026
Supplement Date Received by Manufacturer06/15/2026
Initial Report FDA Received Date03/30/2026
Supplement Report FDA Received Date07/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-