MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS LIFESCIENCES SAPIEN 3 ULTRA RESILIA TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS LIFESCIENCES SAPIEN 3 ULTRA RESILIA TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9755RSL29A
Medical Device Problem Codes Migration or Expulsion of Device (1395); Perivalvular Leak (1457)
Health Effect - Clinical Codes Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Date of Event 03/12/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Per the instructions for use (ifu), paravalvular leak (pvl) is a known potential adverse event associated with bioprosthetic heart valves and the transcatheter valve replacement (tvr) procedure.Paravalvular leak refers to blood flowing through a channel between the structure of the implanted valve and the landing zone, as a result of a lack of appropriate sealing of the valve to the target site.Some pvl is not uncommon post-deployment.Many cases are mild to moderate, and either resolve over time or do not cause symptoms.Others may be more clinically significant and require intervention.The patient screening manual and the procedure didactic identify several procedural and anatomical factors which could contribute to pvl, including device malposition, inaccurate measurement of the landing zone, uneven distribution of calcium on the landing zone, bulky or severe calcification, an elliptical annulus shape, and valve under-sizing.The mechanism behind worsening or late pvl is not well understood but may be related to cardiac remodeling.Edwards extensively trains physicians before they are qualified to use the sapien transcatheter heart valves (all models).The device training manuals instructs the operator on proper imaging screening requirements, including the use of good quality echocardiography and/or computed tomography (ct) to appropriately measure the landing zone, assess the content and distribution of calcium, and other patient anatomical factors.Multiple imaging modalities should be considered during valve size selection.Contraindications, important considerations when assessing the valve, and choosing the proper transcatheter heart valve are also discussed.The device training manuals also instruct the operator on proper positioning and deployment of the valve, including patients screening and procedural considerations.Device preparation, approach, deployment, imaging, procedure-specific training manuals, and proctored procedures are also included.Per the instructions for use (ifu), valve migration requiring intervention is a known potential adverse event associated with transcatheter valve replacement (tvr).According to the literature review, valve migration results when forces acting on the transcatheter heart valve (thv) overcome the strength of attachment of the valve to the annulus.Stent valves are subjected to antegrade ejection forces during systole.Less-than-severe and non-uniformly distributed calcification of the leaflets, incorrect bioprosthetic valve sizing, and incomplete frame expansion can contribute to valve migration.Additionally, residual overhanging leaflets can exert downwards force during diastole, causing migration of the thv towards the ventricle.The device training manuals instruct the operator on proper positioning and deployment of the valve, including all procedural and anatomical considerations.Physicians are extensively trained by edwards before they are qualified to use the sapien transcatheter heart valves (all models).Training includes patient screening, device preparation, approach, deployment, imaging, procedure-specific training manuals, and proctored procedures.The correct alignment and positioning of the device at the point of deployment is emphasized as a key factor in the placement and fixation of the device.Operators are also instructed to use fluoroscopy as the primary method of visualization for positioning and deployment.In patients with high-risk anatomical features for migration (i.E., minimal leaflet calcification), balloon valvuloplasty may indicate potential balloon movement during valve deployment.In this case, there was no allegation or indication that a product malfunction contributed to this adverse event.Investigation results suggest procedural factors may under sized valve may have caused or contributed to the migration.Investigation results suggest migration of the device may have caused or contributed to the paravalvular leak.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.*literature reference for file: philip y.K.Panga, et al.A survivor of late prosthesis migration and rotation following percutaneous transcatheter aortic valve implantation.Eur j cardiothoracic surg (2012).
 
Event or Problem Description
As reported by the field clinical specialist (fcs) approximately 1 year and 1 month post transfemoral transcatheter aortic valve replacement with a 29mm sapien 3 ultra resilia valve the patient was brought back for a balloon aortic valvuloplasty (bav) due to severe paravalvular leak (pvl) and being symptomatic.The patient presented with shortness of breath, hypoxia, and weight gain.The pvl went from severe to none/trace.The patient is in stable condition.There was no aortic insufficiency or pvl after post initial implant.The initial implant was under expanded with some degree of migration into the ventricle.Echo less 16 days prior to bav showed mild to moderate aortic valve regurgitation, a peak velocity of 255 cm/s, mean gradient of 15 mmhg and an aortic valve area of 1.7 cm2.
 
Additional Manufacturer Narrative
A supplemental mdr is being submitted to correct the h6 clinical code.Sections g6 and h2 in this section have been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS LIFESCIENCES SAPIEN 3 ULTRA RESILIA TRANSCATHETER HEART VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key24734331
Report Number2015691-2026-13654
Device Sequence Number10421358
Product Code NPT
UDI-Device Identifier00690103215823
UDI-Public(01)00690103215823(17)260709(11)230710
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P140031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/09/2026
Device Model Number9755RSL29A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2026
Supplement Date Received by Manufacturer05/01/2026
Initial Report FDA Received Date03/30/2026
Supplement Report FDA Received Date05/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
-
-