Model Number 42-9368
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record for lot 30258966 was reviewed and the product was produced according to product specifications.Sample evaluation confirmed the returned tube exhibited a hole just below the connector with associated external denting and internal wall irregularities; however, no manufacturing deficiencies were identified.A root cause could not be determined.All information reasonably known as of 27 mar 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Event or Problem Description
Avanos medical inc.Received a single report that referenced three different incidences, which were associated with separate units, involving two different patients.This is the first of three reports.Refer to 9611594-2026-00153 for the second report refer to 9611594-2026-00154 for the third report.The customer reported ¿hole in the tube after admission¿ occurring during use of a corflo nasogastric feeding tube.The tube required replacement.Per additional information received on 28-aug-2025, the defect was described as a hole located on the yellow tubing next to the purple connector, occurring outside of the patient.No patient injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH ENFIT CONNECTOR
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp 84048
MX   84048
Manufacturer Contact
tamara cardona
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key24735495
Report Number9611594-2026-00152
Device Sequence Number13367343
Product Code KNT
UDI-Device Identifier00350770460925
UDI-Public(01)00350770460925(240)104603302(17)280514(10)30258966
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
K821906
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number42-9368
Device Catalogue Number42-9368
Device Lot Number30258966
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2025
Initial Report FDA Received Date03/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
Patient Weight90 KG
-
-