|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3543257 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Granuloma (1876); Paresthesia (4421)
|
| Date of Event |
10/08/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast prosthesis rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a primary breast augmentation procedure with a mentor memorygel breast implant gel breast prostheses that both ruptured post implantation.Bilateral rupture was diagnosed via mri.Additionally, mri results indicated silicone impregnated internal mammary lymph nodes.As a result, the patient underwent bilateral explantation with no replacement breast prostheses.This medwatch report is for the patient¿s left breast prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2026, the mentor failure analysis lab received the device for evaluation.On (b)(6) 2026, mentor completed its investigation on the suspect medical device.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned device revealed that the breast implant was found to be ruptured.In addition, siltex cracking was noted on the edges of the rupture.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.Granulomas are noncancerous nodules that are a common tissue reaction to the presence of a variety of foreign materials, such as silicone.Paresthesia is an abnormal sensation such as tingling, tickling, pricking, numbness or burning of a person's skin with no apparent physical cause.This includes increased or decreased sensitivity or sensation.The manifestation of a paresthesia may be transient or chronic.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
After clinical/secondary review of this file performed on 31-mar-2026, it was decided to add code paresthesia (e013403), to more accurately capture the reported event.It was reported that the patient felt abnormality underneath the left areola.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|