• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 DPS HD IMP TIP; HIP INSTRUMENT - IMPACTOR TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 DPS HD IMP TIP; HIP INSTRUMENT - IMPACTOR TIP Back to Search Results
Catalog Number 200002302
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11: additional manufacturer narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unspecified surgical procedure it was observed that the head impactor tip broke into three pieces.It was reported that there was a delay in the procedure due to the event.No fragments were generated.There were no reports of injuries, medical intervention or prolonged hospitalization.No further information was provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # - pc-(b)(4) investigation summary - according to the information provided: "it was reported that the item- 200002302- broken into pieces".The product was returned to depuy synthes for evaluation.Visual inspection of the returned device found the distal portion broken.Broken piece was returned for evaluation.No other anomaly was identified on the device surface.A specific root cause cannot be established at this moment.This issue is being addressed by the depuy synthes quality system.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the dps hd imp tip would have contributed to the complained device issue.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through depuy synthes quality system.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot - the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DPS HD IMP TIP
Common Device Name
HIP INSTRUMENT - IMPACTOR TIP
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24740865
Report Number1818910-2026-05485
Device Sequence Number15808049
Product Code HWA
UDI-Device Identifier10603295525875
UDI-Public(01)10603295525875(10)SO2067476
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number200002302
Device Lot NumberSO2067476
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/11/2026
Supplement Date Received by Manufacturer04/07/2026
04/20/2026
Initial Report FDA Received Date03/31/2026
Supplement Report FDA Received Date04/08/2026
04/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/09/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-