| Catalog Number |
J548G |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 3/31/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, the following was obtained: (b)(4).Please provide the following information for each case: ? how many devices demonstrated the alleged deficiency in each procedure? maybe one for each case? when was the needle detached/pulled off from the suture (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify: unk ? please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq) (b)(6).Please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.The device has been received at sukagawa and will be shipped.Please check rmao.Tracking number: (b)(4).D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, a manufacturing record evaluation could not be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an ophthalmic surgery on an unknown date and suture was used.During an unknown ophthalmology, the suture was detached from the needle easily.It was occurred with 2 sutures in a row.It was not a control release needle.Another product was used to complete the case.There were no adverse consequences to the patient.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 4/20/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.Additional information: d9, h3, h6.H3 investigational summary: the product was returned to eth for evaluation.Visual inspection revealed that three detached needles, one needle suture piece and two folder packets were received for analysis.The product code j548g is double armed.Upon inspection of the three detached needles, the closure channel was noted to be as expected.The barrel hole of the needles was examined under magnification, and no suture remnant was noted.However, one needle was found bent, likely due to used.In addition, the needle suture was visually inspected, and the attachment area was noted to be as expected.The suture was examined and the end was noted to be cut, likely by a surgical instrument.The other section of the suture and one suture were not returned for evaluation.As the suture is absorbable and the duration and conditions of environmental exposure could not be determined, functional testing was not performed.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Given the current condition of the returned sample, no conclusion could be reached as to what caused the reported complaint.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|