SMITH & NEPHEW, INC. FST FIX FLX REV CRVD INSRTR BNDR CANN ST; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
|
Back to Search Results |
|
| Catalog Number |
72205325 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a meniscus repair surgery, the t1 implant of one (1) fast-fix flex deployed successfully; however, when the deployment knob was depressed, the t2 implant did not anchor securely and came off.The procedure was resumed, after a 5 minute delay, using an equivalent s+n back-up.Additional anesthesia was not required as consequence of the surgical prolongation.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h3, h6: the reported device was received for evaluation.A visual inspection revealed the device was returned in original packaging with the batch number of the complaint on the label.The t1 was returned off the needle but attached to the suture.The t2 and suture were returned on the needle.The insertion device has no visible defect.Bio debris is present.A functional evaluation was performed on the returned device and found the actuator cycled the t2 off the needle and continued to cycle as intended.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (failure to fully actuate the device behind the meniscus during implantation).Based on this investigation, the need for corrective action is not indicated.
|
| |
|