Additional information: d9, g3, h2, h3, h6, h11.One 8.5f swartz braided introducer sheath was received for evaluation.Functional testing determined a leak was noted in the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.A puncture was noted in the proximal and distal seals.The batch record was reviewed to ensure that each manufacturing and inspection operation was performed.The ifu states: do not remove dilator or catheter rapidly.Damage to the valve may occur, potentially compromising hemostasis.
|