• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. RHINO-LARYNGOFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. RHINO-LARYNGOFIBERSCOPE Back to Search Results
Model Number ENF-P4
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.Based on the results of the investigation, it is likely the following led to the malfunction: a stress problem was identified.The most probable cause is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the rhino-laryngofiberscope exhibited the insertion part connecting tube coating peeling off.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RHINO-LARYNGOFIBERSCOPE
Common Device Name
RHINO-LARYNGOFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24744451
Report Number3002808148-2026-08932
Device Sequence Number15728630
Product Code EOB
UDI-Device Identifier04953170411069
UDI-Public04953170411069
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
K011869
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberENF-P4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/12/2026
Initial Report FDA Received Date03/31/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-