• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS LIFESCIENCES SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS LIFESCIENCES SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9750TFX20
Medical Device Problem Codes Perivalvular Leak (1457); Malposition of Device (2616)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/03/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is ongoing.
 
Event or Problem Description
As reported by our german affiliate during a transfemoral transcatheter aortic valve replacement (tavr) with a 20mm sapien 3 ultra valve, although the valve was appropriately positioned between the balloon markers prior to deployment, during valve deployment the 20mm commander delivery system (ds) balloon inflated unexpectedly, with the proximal (outflow) portion inflating first while the distal (inflow) portion remained snared with the valve and then abruptly inflating with the proximal balloon part, causing an unpredictable valve deployment.Consequently, the valve was implanted too high in the aortic position, with a higher grade of paravalvular leak observed on final fluoroscopy.The delivery system was subsequently removed without any issues, and no damage was observed on the 14fr esheath+ upon withdrawal.A second valve was then implanted more ventricular which resulted in no paravalvular leak.The procedure was therefore completed successfully.No injury was reported, and the patient was discharged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS LIFESCIENCES SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key24744942
Report Number2015691-2026-13711
Device Sequence Number9101271
Product Code NPT
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P140031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number9750TFX20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2026
Supplement Date Received by Manufacturer05/07/2026
Initial Report FDA Received Date03/31/2026
Supplement Report FDA Received Date05/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-