BECKMAN COULTER, INC. AU5800 CHEMISTRY ANALYZER; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
|
Back to Search Results |
|
| Model Number |
AU5800-20 CHEMISTRY ANALYZER AU5800 AUTOMATION READY |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the generation of erratic ion-selective electrolytes (ise) patient results, including sodium (na), chloride (cl) and potassium (k), on their au5800 clinical chemistry analyzer.There was no report of change to treatment, injury, or death associated with event.The customer did not provide patient data or demographics.
|
| |
|
Additional Manufacturer Narrative
|
|
Beckman coulter field service engineer (fse) evaluated the au5800 chemistry analyzer and identified the probable cause as a defective degasser.The fse replaced degasser, tubing and degasser pump to resolve the issue.Section a2, a3, a4, a5 and a6: information not provided by customer.The beckman coulter internal identifier is (b)(4).
|
| |
|