Catalog Number 60PFSS30
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filled as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was stated that the tracheostomy tubes broke off at the site where the ventilator circuit was connected.It was a very dangerous situation because it suddenly disconnected the child from the ventilator, requiring the parent or caregiver to quickly remove the broken trach, insert a new one, and reconnect the ventilator.There were patient involvement and no reported patient harm.
 
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product was returned; therefore, no visual and functional tests were performed, the reported complaint could not be confirmed, and the root cause could not be determined.If the product is returned, the manufacturer will reopen this complaint for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA FLEXTEND PEDIATRIC AND NEONATAL TRACH TUBE
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
ICU MEDICAL HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana 22425
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key24746037
Report Number3012307300-2026-03099
Device Sequence Number18319827
Product Code JOH
UDI-Device Identifier15021312516760
UDI-Public(01)15021312516760(17)300601(10)6115613
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K944178
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number60PFSS30
Device Lot Number6115613
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2026
Supplement Date Received by Manufacturer04/02/2026
Initial Report FDA Received Date03/31/2026
Supplement Report FDA Received Date04/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-