| Catalog Number |
TSVMH10 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Abscess (1690); Unspecified Infection (1930); Osteolysis (2377)
|
| Date of Event |
01/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A4: weight unknown / not provided.A summary investigation has been completed for bone loss and infection events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss and infection have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant at tooth site #11 was removed due to a lack of stability associated with infection.Examination and x-ray imaging confirmed that the implant site was infected.This was noted during an implant uncovering procedure.The implant removal tool was broken during removal.An additional appointment was required to replace the implant.Symptoms as a result of the event: abscess and pain.
|
| |
|
Search Alerts/Recalls
|
|
|