Model Number FG11815
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation (2001)
Date of Event 03/03/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a perforation was identified requiring additional intervention.An enroute transcarotid stent was selected for use in a right sided transcarotid artery revascularization (tcar) procedure.During the procedure, the stent was placed, post dilated, and then an angiogram was performed.The angiogram revealed a perforation in the internal carotid artery (ica).The physician elected to convert the procedure to a carotid endarterectomy (cea) and the stent was explanted.
 
Additional Manufacturer Narrative
Detailed product information was not provided to bsc.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other specific product information.If additional details become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that a perforation was identified requiring additional intervention.An enroute transcarotid stent was selected for use in a right sided transcarotid artery revascularization (tcar) procedure.During the procedure, the stent was placed, post dilated, and then an angiogram was performed.The angiogram revealed a perforation in the internal carotid artery (ica).The physician elected to convert the procedure to a carotid endarterectomy (cea) and the stent was explanted.
 
Additional Manufacturer Narrative
Detailed product information was not provided to bsc.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other specific product information.If additional details become available, a supplemental report will be submitted.Investigation results: ship history review: a ship history review for the reported upn was performed for the reporting customer.Batches 18448262 and 184261153 were most recently shipped to the reporting facility in the 3-year period preceding the procedure date.There were no manufacturing deviations for any of the batch numbers that could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review: a review of the enroute transcarotid stent device confirmed that the reported event is a known event defined in the product's risk management documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENROUTE TRANSCAROTID STENT
Common Device Name
STENT, CAROTID
Manufacturer (Section D)
SILK ROAD MEDICAL
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer (Section G)
SILK ROAD MEDICAL INC.
27th ave north plymouth
minneapolis MN 55447
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key24747182
Report Number2124215-2026-17689
Device Sequence Number12954950
Product Code NIM
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberFG11815
Device Catalogue NumberFG11815
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2026
Supplement Date Received by Manufacturer06/03/2026
Initial Report FDA Received Date03/31/2026
Supplement Report FDA Received Date07/02/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexFemale
-
-