MEDOS INTERNATIONAL SÃ RL HD LAPSCP,EP,5.5,30,300; LAPAROSCOPE, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
| Catalog Number |
242633 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: - other damages caused by customer.- outer tube damaged, distal tip distal tip damaged scratches and knicks.¿ illumination distal tip fiber damaged.· particulate, optics particulate under distal lens particulate in system.- optical system, optical components prism loose broken lenses in optical system.- cosmetic issue scratches/dents labeling : illegible etch, visual : deformed/bent, broken (2+ pieces) : chipped/shaved/delaminated, visual : scratched/nicked.Per service report, this complaint was confirmed.Based on the investigation findings, the potential root cause is traced to user error.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review a manufacturing record evaluation was performed for the finished device, and no non-conformance was identified.
|
| |
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the hd lapscp,ep,5.5,30,300 device was broken (2+ pieces) : chipped/shaved/delaminated.It was reported in the service order that the device had an undetermined malfunction.It was unknown when the event occurred.There were no reports of delays to a surgical procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2026.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|
|
|