(b)(4) date sent: 3/31/2026 b3: exact event date unk, entered 1/1/2026 as only the year was provided.D4: batch # 313d90.Investigation summary the product was returned to ethicon endo-surgery for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample determined that the gcs40g device was received with no apparent damage, with a reload present and two loose stapler retainers.The reload was received fully loaded with staples, the drivers and knife recessed below the cartridge deck and the washer returned detached.The device was tested for functionality with a test reload and it fired, cut, and formed the staples as intended.The cut line and the staple line were completed and the staples meet the staple form release criteria.It should be noted that to reload the device insert the new reload into the metal housing and snap it into position.The tracks on each side of the reload should be used as guides to align the reload within the jaws of the instrument.When the reload is properly aligned, push the reload into the instrument until it is fully seated.Remove the staple retainer.Check that the reload is held firmly within the jaws.If the reload is removed from the device, whether the reload is spent or not, and if the staple retainer has already been removed from the reload, the reload cannot be reloaded into the device.Please reference the instructions for use for additional information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Device history review a manufacturing record evaluation was performed for the finished device batch number 313d90, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|