|
(b)(4).Date sent: 3/31/2026.D4: batch # 765d24.Investigation summary: the product was returned to ethicon endo-surgery for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample determined that the gcs40g device was received with no apparent damage and with a reload loaded in the device.The reload was received fully loaded with staples, with the washer uncut, with the drivers and with the knife recess below the reload deck.The reload was not properly loaded in the device, as the retaining pin was not connected to the coupler and pushrod.These facts indicate that the reload was removed and reinserted incorrectly and/or incompletely after the retainer was removed.The returned reload was pulled out of the device and placed back on; the device opened and closed without any difficulties.The device was tested for functionality with an returned reload and the device fired, forming all staples and cutting as intended.The cut line and the staple line were completed and the staples meet the staple form release criteria.In addition, a tyvek returned along with the device it should be noted that if the reload is removed from the device, whether the reload is spent or not, and if the staple retainer has already been removed from the reload, the reload cannot be reloaded into the device.Please reference the instructions for use for the complete guide loading and reloading the instrument.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Device history review: a manufacturing record evaluation was performed for the finished device batch number 765d24, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
It was reported that during exploratory laparotomy sigmoidectomy procedure the device was used once, and when a reload is placed, the stapler engages with the load, but it does not allow you to fire it or engage the handle.There were no patient adverse event.
|