Catalog Number 3504254BC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 09/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient implanted with two 425cc mentor memorygel breast implant experienced bilateral ruptures post-operatively.As a result, the patient underwent explantation on (b)(6) 2026.
 
Additional Manufacturer Narrative
Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection, microscopic examination, and shell thickness of the returned device.Visual analysis of the returned device revealed that the breast implant was found to be ruptured from both aspects.Additionally, it was received in two segments.Microscopic examination was performed, and the cause of the tear could not be identified.Therefore, a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation but is more likely to occur the longer the implant is implanted.The following may cause implant to rupture: stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.Asymmetry may be attributed to one or more of the following: contracture of the fibrous capsule, seroma or hematoma, development of postoperative breast dysplasia, unilateral discrepancy in muscle development, deflation of the implant, incorrect choice of implant shape or size, and surgical technique.Asymmetry is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key24748529
Report Number1645337-2026-03720
Device Sequence Number21105137
Product Code FTR
UDI-Device Identifier00081317000518
UDI-Public(01)00081317000518(17)200410(10)6924048
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/10/2020
Device Catalogue Number3504254BC
Device Lot Number6924048
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/11/2026
Supplement Date Received by Manufacturer04/13/2026
Initial Report FDA Received Date03/31/2026
Supplement Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/10/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-