|
H11: additional manufacturer narrative: bioprosthetic tissue valves can deteriorate with time and eventually fail contributing to regurgitation and/or stenosis.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular regurgitation and/or stenosis.In this case, minimal information regarding this valve explant procedure was received and attempts to obtain additional information regarding the condition of the device, patient's medical history, or possible comorbidities have been made.The root cause of this event cannot be determined with the available information.However, this event was most likely impacted by the progression of the patient's underlying valvular disease pathology with or without structural valve deterioration and/or nonstructural dysfunction.The subject device is not available for evaluation as attempts were made to retrieve the device, but no device was returned.The device history record (dhr) was not reviewed as no information regarding a device failure mode was provided.Therefore, a review of manufacturing and/or device component issues that may have contributed to the reported event is unable to be performed.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.If new information becomes available, a supplemental report will be submitted.
|
|
It was learned through implant patient registry that a patient with a 29mm 3300tfx aortic valve was explanted after an implant duration of 9 years, 4 months due to unknown reasons.The explanted valve was replaced with a 27mm 11060a valved conduit.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'.The information reported may or may not be related to the edwards device.
|