| Catalog Number |
LH-0029 |
| Medical Device Problem Codes |
Material Twisted/Bent (2981); Activation Problem (4042)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.The device has not been returned to the manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that after administering the medication, when health professional tried to flush the line and remove the needle, the safety mechanism is not functioning properly and failed to engage.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
Additional information received 01 april 2026: no needlestick accidents.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of difficulty advancing the safety mechanism was confirmed to be caused by a bent needle.However, the exact cause of the bent needle remains unknown.The product returned for evaluation was one 22g safestep infusion set w/o y-site.The returned product sample was evaluated and the needle was observed to be bent.During evaluation the safety mechanism was disengaged to inspect the needle tip.The following observations were noted during the sample evaluation: usage residue was seen on the sample which proved that the product had experienced at least some use.The bend in the needle had occurred near the needle base.The tip of the needle bevel appeared unremarkable.An attempt to advance the safety was successful; however, resistance was encountered when passing the safety over the bent region of the needle.Based on the evidence provided with the returned sample it is unknown when or how the bend occurred in the needle.Damage to the needle could have occurred at the manufacturing facility, during shipping, during use, or during storage of the product, and no evidence was observed which supported a specific root cause of the event.This complaint will be recorded for future trending and monitoring purposes.
|
| |
|