| Model Number |
MAJ-2315 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that during a therapeutic ercp procedure, the distal cover dislodged.The distal cover was removed as a foreign body from the patient's airway.The patient's condition is reported as stable.Further follow-up had been conducted and is currently pending for additional information.This report is 2 of 2.
|
| |
|
Event or Problem Description
|
|
Additional information received identified the procedure was a endoscopic retrograde cholangiopancreatography for investigation of stones in the bile duct and removal of stents.The distal cap had to be removed as a foreign body from the airway requiring airway support.Additionally the cover was dislodged in the mouth or throat and removed manually or via the endoscope using grasping forceps.The procedure was completed, but with a 10-minute delay and extension of sedation time, requiring re-insertion of the scope to locate and retrieve the missing distal cover and to stabilize the airway.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the approved final investigation.Updated fields- b5, h2, h3, h4, h6, h7, h9 and h11.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined and the event could not be confirmed.However, it is likely cause of failure was due to the following mechanism.The distal end cover was not properly attached and dropped off due to the load during the procedure.Since it is difficult to visually detect the state of attachment of this product, it is feasible that the attachment of the distal cover was insufficient.Olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted for a correction to the b5 event details.
|
| |
|
Event or Problem Description
|
|
Follow-up received indicated that no airway support was needed.The distal cap fell into the patient's mouth and throat.The endoscope was reinserted to find the missing cap.The procedure time was extended by 10 minutes.
|
| |
|