SMITH & NEPHEW, INC. ULTRA FAST-FIX ASSEMBLY - CURVED; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
|
Back to Search Results |
|
| Catalog Number |
72201491 |
| Medical Device Problem Code |
Firing Problem (4011)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during a meniscus repair, both t implants of one (1) ultra fast-fix curved fell off at the same time as the first deployment.The procedure was resumed, using an equivalent s+n backup; however, it is unknown if there was a surgical delay.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alter the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a meniscus repair, both t implants of one (1) ultra fast-fix curved fell off at the same time as the first deployment.The procedure was resumed, using an equivalent s+n backup, with a non-significant delay.No additional anesthesia was required.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h3, h6: part of the device, used in treatment, was received for evaluation.A visual evaluation showed the device was returned in original packaging with the batch number of the complaint on the label.The t1 and t2 were not returned.A partial suture was returned.The variable depth straw was not returned.Bio debris is present.A functional evaluation could not be performed because both implants have already been deployed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Factors that can contribute to the reported event include unintended second actuation of the device or excessive retraction of the device causing t2 to be pulled from the needle.Based on this investigation, the need for corrective action is not indicated.
|
| |
|