Model Number VSMP15
Medical Device Problem Code Fire (1245)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a breast procedure, the devices were involved in a brief open flame at the distal tip of the electrode.The procedure was momentarily paused due to the sentinel event involving the open flame.The technique used involved a metal pointed clamping instrument clamped onto tissue, with the electrode placed touching the jaws of the instrument and activated, causing the electrical circuit to move through the jaws to the target tissue.After the electrode was removed from the instrument, a flame briefly appeared at the distal tip and then went away quickly.A new generator was retrieved and the case resumed after replacing the electrode and opening an older smoke evac pencil from another manufacturer.There was no patient injury.
 
Additional Manufacturer Narrative
D10 concomitant products: vlft10gen, vlft10gen ft series energy platformx1 (serial#:unknown), e1455, e1455 the edge ent/ima electrode x50 (lot#:unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a breast procedure, the devices were involved in a brief open flame at the distal tip of the electrode.The procedure was momentarily paused due to the sentinel event involving the open flame which was extinguished by stopped pressing the activating energy button.The technique used involved a metal pointed clamping instrument clamped onto tissue, with the electrode placed touching the jaws of the instrument and activated, causing the electrical circuit to move through the jaws to the target tissue.After the electrode was removed from the instrument, a flame briefly appeared at the distal tip and then went away quickly.There was no operating room fire.A new generator was retrieved, and the case resumed after replacing the electrode and opening an older smoke evac pencil from another manufacturer.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: b5, g3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB SM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24751610
Report Number2182208-2026-15402
Device Sequence Number21368605
Product Code GEI
UDI-Device Identifier10884521836365
UDI-Public(01)10884521836365(17)290630(10)2503211X
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K240572
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVSMP15
Device Catalogue NumberVSMP15
Device Lot Number2503211X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2026
Supplement Date Received by Manufacturer03/24/2026
Initial Report FDA Received Date04/01/2026
Supplement Report FDA Received Date04/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Patient SexFemale
-
-