RESPIRONICS, INC. GARBIN EVO, LINDE; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number LD2110X23B
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported failure of vent inop indicating that the therapy cpu remained in boot monitor software is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
 
Event or Problem Description
The manufacturer received a report that a garbin evo ventilator was returned for maintenance service.There were no reports or allegations of patient harm or injury.The device was not reported to be in patient use.During the evaluation of the device at a third-party service center, the device log files were successfully downloaded and reviewed in full.A ¿vent inop¿ alarm was identified, indicating that the therapy cpu remained in boot monitor software.The available service documentation did not specify that any components required replacement to resolve the issue.No additional actionable error codes were identified.Due to the 4-year preventive maintenance, the air inlet foam filter, particulate filter and muffler were replaced.The customer complaint was confirmed.The device has been serviced, inspected, tested, and met acceptance criteria in accordance with all of the applicable customer requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GARBIN EVO, LINDE
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key24752671
Report Number2518422-2026-010267
Device Sequence Number10360205
Product Code CBK
UDI-Device Identifier00606959057432
UDI-Public(01)00606959057432
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberLD2110X23B
Device Catalogue NumberLD2110X23B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2026
Initial Report FDA Received Date04/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-