| Catalog Number |
72290003 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament reconstruction surgery, the ultrabutton adjustable loop suture fracture and could not be used.The procedure was resumed, without any delay, using an equivalent s+n back-up.No further complications were reported.The patient¿s condition is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This event was revisited and re-evaluated.Based on the assessment, the broken section of the suture does not pose a risk to the patient.Loop suture breakage may result in a short procedural delay and/or require use of a backup device to continue procedure but is not likely to cause or contribute to a death or serious injury; therefore, it was determined that this case does not meet the threshold for reporting and is classified as a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
|
| |
|
Search Alerts/Recalls
|