|
The tsh reagent lot number was 891754.The ft4 reagent lot number was 883869.The tsh and the ft4 reagents' expiration dates were not provided.The field service engineer (fse) found an issue with the sample probe tubing.He replaced the pinch valve tubings, the sample probe tubings, and primed the sample/reagent probe.He also primed the sipper, adjusted the sample probe liquid level detection voltage, and performed maintenance on the instrument.Qc was performed, and it was within specifications.The fse verified that the instrument was performing within specifications.No further issues were reported after the service visit.The cause of the event was due to a component failure (an issue with the sample probe tubing).
|