|
According to the complaint description the ¿flow sensor calibration test failed multiple times via hamilton mr1 serial number (b)(6).Doctor changed breathing set pipes a few times.When changing the hospitals device for a medidyne loaner but using the same breathing set pipes that were in the hospital, the same flow sensor problem also happened on the medidyne device.¿ although patient involvement was reported, no medical intervention, no health consequences, and no impact were reported for the patient, the user, or any third party.
|
|
Hamilton medical reference number: (b)(4) country of event occurrence is finland.Hamilton coaxial breathing circuit family devices in this product line are classified under product code bzo.¿set, tubing and support, ventilator (with harness).Part number 260167 is listed in the same product family per the manufacturer¿s ifu, and therefore falls under the same fda product classification.This report contains the investigation report as well.The reported issue involved repeated failures of the flow sensor calibration tests on a hamilton-mr1 ventilator (serial number (b)(6).No logfiles or any other evidence were provided to confirm the reported issue.Although patient involvement was reported, no medical intervention, no health consequences, and no impact were reported for the patient, the user, or any third party.According to the instructions for use (ifu) of the breathing circuit set, coaxial, all pre-operational checks, including flow sensor calibration - must be successfully completed before connecting a patient to prevent possible injury, and the user must ensure that an alternative ventilation source is available.The ifu further specifies that a product must be discarded if it fails two subsequent pre-operational checks or shows any sign of damage.In this case, the hospital¿s breathing circuit failed multiple calibration attempts.A device that has not passed calibration will continue to prompt completion of the required pre-operational checks in accordance with the ifu, and under these conditions the ventilator is not intended for patient connection.This calibration ensures that the ventilator can accurately measure the airflow and airway pressure being delivered to the patient and does not immediately indicate that the ventilator is unsafe.It is a diagnostic check to confirm that the sensor is functioning as expected.Devices always need to undergo self-test and calibration before usage.If the flow sensor calibration failure alarm is ignored, the device would continuously alarm when used.If the test is not performed at all (which is against instruction in ifu) there is a potential that the measurement/values shown are not fully accurate.In conclusion, a failed flow sensor calibration should not be viewed as a malfunction or a critical failure, but as part of the preventive safety and maintenance measure.As long as the underlying issue is addressed properly, the ventilator can be safely used on patients.The investigation showed that the repeated calibration failures followed the breathing circuit rather than the ventilator itself.The root cause is therefore consistent with a faulty breathing circuit leading to inaccurate flow measurement during pre-use testing.Replacing the breathing circuit resolved the issue, and it was confirmed that the calibration tests have passed.The investigation is therefore considered complete and the case closed.
|