| Catalog Number |
0500320E |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
03/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The registered nurse (rn) reported to fresenius that a dialyzer blood leak occurred during the patient's hemodialysis (hd) treatment.Additional information was obtained during follow-up with the clinical manager (cm).The reported issue occurred upon treatment initiation.The fresenius 2008t machine alarmed as expected with a blood leak alarm.Blood leak test strips were not used to test for the presence of blood.The patient care technician (pct) visually observed pink dialysate and immediately clamped the venous and arterial lines.The patient's estimated blood loss (ebl) is approximately 250 ml.No serious injury or required medical intervention resulted from the reported issue.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be discarded and unavailable to be returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
|
| |
|
Event or Problem Description
|
|
The registered nurse (rn) reported to fresenius that a dialyzer blood leak occurred during the patient's hemodialysis (hd) treatment.Additional information was obtained during follow-up with the clinical manager (cm).The reported issue occurred upon treatment initiation.The fresenius 2008t machine alarmed as expected with a blood leak alarm.Blood leak test strips were not used to test for the presence of blood.The patient care technician (pct) visually observed pink dialysate and immediately clamped the venous and arterial lines.The patient's estimated blood loss (ebl) is approximately 250 ml.No serious injury or required medical intervention resulted from the reported issue.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be discarded and unavailable to be returned to the manufacturer for evaluation.
|
| |
|