• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OGDEN MANUFACTURING PLANT OPTIFLUX F200NRE HF 12/CS 1.9SA EBEAM; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OGDEN MANUFACTURING PLANT OPTIFLUX F200NRE HF 12/CS 1.9SA EBEAM; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 0500320E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 03/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
The registered nurse (rn) reported to fresenius that a dialyzer blood leak occurred during the patient's hemodialysis (hd) treatment.Additional information was obtained during follow-up with the clinical manager (cm).The reported issue occurred upon treatment initiation.The fresenius 2008t machine alarmed as expected with a blood leak alarm.Blood leak test strips were not used to test for the presence of blood.The patient care technician (pct) visually observed pink dialysate and immediately clamped the venous and arterial lines.The patient's estimated blood loss (ebl) is approximately 250 ml.No serious injury or required medical intervention resulted from the reported issue.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be discarded and unavailable to be returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Additional Manufacturer Narrative
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
 
Event or Problem Description
The registered nurse (rn) reported to fresenius that a dialyzer blood leak occurred during the patient's hemodialysis (hd) treatment.Additional information was obtained during follow-up with the clinical manager (cm).The reported issue occurred upon treatment initiation.The fresenius 2008t machine alarmed as expected with a blood leak alarm.Blood leak test strips were not used to test for the presence of blood.The patient care technician (pct) visually observed pink dialysate and immediately clamped the venous and arterial lines.The patient's estimated blood loss (ebl) is approximately 250 ml.No serious injury or required medical intervention resulted from the reported issue.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be discarded and unavailable to be returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX F200NRE HF 12/CS 1.9SA EBEAM
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key24756129
Report Number0001713747-2026-00144
Device Sequence Number9138181
Product Code KDI
UDI-Device Identifier00840861100163
UDI-Public00840861100163
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K162488
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0500320E
Device Lot Number25SU04012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2026
Supplement Date Received by Manufacturer04/09/2026
Initial Report FDA Received Date04/01/2026
Supplement Report FDA Received Date04/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS 2008T MACHINE; FRESENIUS 2008T MACHINE; FRESENIUS BLOODLINES; FRESENIUS BLOODLINES
Patient Age54 YR
Patient SexMale
Patient Weight127 KG
-
-