U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ESC PINNACLE; EXTRACTION INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ESC PINNACLE; EXTRACTION INSTRUMENTS Back to Search Results
Catalog Number 221800010
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the tip of liner removal tool was broken off.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information provided: ""it was reported that the tip of liner removal tool was broken off".The product was returned to depuy synthes for evaluation.Visual analysis of the returned device found its tip broken and deformed.Broken fragment was not returned for evaluation.Device had signs of usage on its overall surface and no additional damage was identified on it.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces such as prying out the liner instead of tapping to gently breaking the bond between the tapers.Properly handling and¿attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the pinn esc liner extractor tip would have contributed to a device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESC PINNACLE
Common Device Name
EXTRACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24756140
Report Number1818910-2026-05648
Device Sequence Number13052801
Product Code HWB
UDI-Device Identifier10603295099505
UDI-Public(01)10603295099505(10)NW137169
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number221800010
Device Lot NumberNW137169
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/15/2026
Supplement Date Received by Manufacturer05/06/2026
Initial Report FDA Received Date04/01/2026
Supplement Report FDA Received Date05/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/01/2001
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexFemale
-
-