There was an allegation of questionable results with multiple assays for a patient sample tested on the cobas integra 400 plus.This medwatch will apply to the hdl assay.Please refer to the medwatch with a1.Patient identifier (b)(6) for information related to the lipase assay.Please refer to the medwatch with a1.Patient identifier (b)(6) for information related to the cholesterol assay.Please refer to the medwatch with a1.Patient identifier (b)(6) for information related to the ldl assay.Please refer to the medwatch with a1.Patient identifier (b)(6) for information related to the triglycerides assay.Initial results: lipase: 8,859 u/l, cholesterol: 4,074 mg/dl, hdl: 1,611 mg/dl, ldl: 2,077 mg/dl, triglycerides: 3,133 mg/dl.Repeat results: lipase: 4,678 u/l, cholesterol: 167 mg/dl, hdl: 69 mg/dl, ldl: 88 mg/dl, triglycerides: 133 mg/dl.Repeat results after a 1:10 manual dilution: cholesterol: 196 mg/dl, hdl: 80 mg/dl, ldl: 101 mg/dl, triglycerides: 154 mg/dl.
|