| Catalog Number |
71631140 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case (b)(4).H11: lot numbers 08km04375 and 23am00380 were received for guide bolt wrenchs reported.However, it is unknown which belongs to this event (case-(b)(4).For lot number 08km04375: manufactured date: 06-oct-2008 / expiration date: 21-feb-2036.For lot number 23am00380: manufactured date: 03-jan-2023 / expiration date: 20-may-2050.
|
| |
|
Event or Problem Description
|
|
It was reported that during an internal fixation surgery, when attempting to remove the drill guide handle from the intertan 11.5mm x 18cm 125d, the tip of a guide bolt wrench broke off inside the drill guide handle (case-(b)(4).Afterwards, another attempt was made to remove the intertan 11.5mm x 18cm 125d from the drill guide handle on the back table, and another guide bolt wrench broke.It is unknown how was the procedure completed or if there was any surgical delay.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.The visual inspection revealed that the device returned heavily worn from use, with the distal tip fractured off with no material returned.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed part number over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|