It was reported that the implantable cardioverter defibrillator exhibited 96 percent inappropriate mode switch caused by oversensing.The device was explanted and replaced.There were no patient consequences.
Additional Manufacturer Narrative
The reported events of no lv output, failure to sense, under-sensing, and inability to program were not confirmed.The device voltage was at elective replacement indicator (eri) upon receipt.Analysis of device image was performed, and no anomalies were noted.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Electrical testing was performed, and no anomalies were noted.