|
W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
PXC201000 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Clinical Endoleak (4843)
|
| Date of Event |
03/04/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2010, a patient underwent an endovascular procedure to treat an abdominal aortic aneurysm using gore® excluder® aaa endoprosthesis devices.Two contralateral leg components (both pxc181000) were implanted bilaterally, and a contralateral component (pxc201000) was added to extend the right leg.The specific device serial number implanted at end of the left leg could not be determined.On an unknown date, type ib endoleaks were observed in both legs.On (b)(6) 2026, a reintervention was performed.Reportedly, at the time of the reintervention, the right internal iliac artery was found to be occluded.The left internal iliac artery was coil embolized.Both legs were extended with gore® excluder® aaa endoprosthesis devices into the external iliac arteries.The endoleaks were resolved, and the patient tolerated the procedure.
|
| |
|
Search Alerts/Recalls
|
|
|